MGA Luxx

MGA Luxx

Regular price$0.00
/

Ideal for long time measurement of industrial combustion, large boilersgas engines and turbinesfurnaces and many more

Basic Equipment

Employs a stable offset measuring technique with NDIR sensors for incredible accuracy and long term stability.

  • Simultaneous analysis of up to 9 gases.
  • NDIR: CO, CO2, NO, NO2, SO2, CH4, C3H8
  • Electrochemical: H2S, H2, O2 (EC or paramagnetic sensor)
  • 7" high resolution TFT color display with swipe and interface
  • Intuitive icons and country specific languages
  • Rugged, aluminum enclosure with rubber molded impact protection ready for harsh industrial environments
  • Advanced algorithms assure quality measurements and high performance operations
  • large internal memory plus external USB drive
  • Functionality to review a measurement session in graphical form on the display and/or send as CSV or PDF reports.
  • Outputs include ethernet, WLAN, USB, RS485 Modbus, 8 ch./4 ch. analog input/output.
  • Integrated sample gas conditioning. Dual gas conditioner (peltier) with dual draining pumps maintain a constant 41 degree (F) dew point
  • Condensate monitoring and alarm on display
Technical Data
Gas measurement Method Meas. range

 

(0…min / max)

Resolution Accuracy **
O2 – Oxygen ECS 0 to 25.00 % 0.01 % 0.2 %
O2 – Oxygen PM 0 to 25.00 % 0.01 % 0.1 %
CO2 – Carbon monoxide NDIR 0 to 5 / 40% 0.01 % ± 0.2 % or 2 % reading
CO – Carbon dioxide NDIR 0 to 1,000 / 10,000 ppm 1 ppm ± 20 ppm or 2 % reading
HC – Hydrocarbons as CH4 NDIR 0 to 1,000 / 10,000 ppm 1 ppm ± 20 ppm or 2 % reading
HC – Hydrocarbons as C3H8 NDIR 0 to 1,000 / 10,000 ppm 1 ppm ± 10 ppm or 2 % reading
NO – Nitric oxide NDIR 0 to 1,000 / 4,000 ppm 1 ppm ± 20 ppm or 2 % reading
NO2 – Nitrogen dioxide NDIR 0 to 500 / 1,000 ppm 1 ppm ± 10 ppm or 2 % reading
SO2 – Sulfur dioxide NDIR 0 to 1,000 / 4,000 ppm 1 ppm ± 10 ppm or 2 % reading
HC – Hydrocarbons as CH4 NDIR 0 to 1,000 / 4,000 ppm 1 ppm ± 20 ppm or 2 % reading
H2S – Hydrogen sulfide ECS 0 to 2,000 / 5,000 ppm 1 ppm ± 10 ppm or 2 % reading
H2 – Hydrogen ECS 0 to 1,000 / 2,000 ppm 1 ppm ± 5 ppm or 2 % reading
OTHER MEASUREMENTS / CALCULATIONS Method Measuring range
Tgas – stack gas temperature NiCrNi 32°F to 2.192°F K-type
Tair – combustion air temperature NiCrNi 32°F to 932°F K-type
Tamb – ambient air temperature RTD 32°F to 212°F PT2000
Diff.pres. – differential pressure Piezoresistive -48 to +48 inH2O
GasFlow – flow velocity – differential pressure Pitot 3 … 100 m/s
AUX – auxiliary connector –

 

standardized external signal

Software for K-thermocouple, 0 … 10 Vdc , 4 … 20 mA, RS485
Combustion calculations – fuel type dependent Software heat loss, ExcAir, Air Ratio (λ), dewpoint
Emission calculations Software mg/Nm3 , reference to O2 , g/kWh, g/s , kg/h
GENERAL TECHNICAL DATA  
Operating system LINUX
Display, operation 7“ TFT (800 x 480 px) colour display, backlit, with touch pad
Data storage type /capacity internal 10.000 data sets, external USB stick
Interface to external PC/notebook Ethernet, Bluetooth, WLAN, RS485
Cable communication interface RS485, RJ45 (Ethernet)
Wireless communication WLAN
Thermal printer external only
Analog output 4 – 20 mA/analog input 4 – 20 mA 8 channel out / 4 channel in/user configurable
Universal analog input – AUX – 0…10 Vdc / 4…20 mA / NiCrNi / RS485
System warming up time 30 minutes (typical)
Warming up temperature NDIR bench 131°F (55 °C)
Mains free operation time / stand-by only 1 hour
Internal battery Li-Ion , 96W, for standby
Operating conditions 41°F to 113°F, RH up to 95% non condensing
Storage temperature -4°F to 122°F
Power supply / consumption 86 to 265 Vac / 47…63 Hz / 105 W (analyzer only)
Enclosure material aluminum, rubber molded impact protection
Protection class IP20 (or IP42 inside transport case)
Dimensions 16.92“ x 11.41“ x 5.9“ (WxHxD)
Weight from 16.5 lbs. for minimal configuration

* overload range of ECS is usable only for short duration

** which ever is larger