The SWG 100 CEM is a stationary analyzer meant for continuous emission monitoring. It works best for industrial applications using the extractive method.
Basic Equipment
- Field replaceable plug play pre-calibrated sensors
- Compact industrial design for up to 6 gas simultaneous measurements
- Uses low cost, reliable electrochemical cells for O2, CO, NO, NO2, SO2 and an infrared module (NDIR) for CO2 measurement
- Advanced sample gas preparation for fast and reliable measurements
- Flexible platform can be used for various combustion applications
- Direct and continuous/discontinuous measurement, with pressure and temperature compensation of all main flue gas parameters
- External measurements (temperature, pressure, etc.) by reading of ext. standard signal
- Simple installation, ready to run delivery and easy to maintain
Technical Data
Measuring ranges |
O2 – Oxygen |
0 … 25 % |
CO2 – Carbon dioxide |
0 … 40 % |
CO – Carbon monoxide |
0 … 10,000 ppm |
NO – Nitric oxide |
0 … 4,000 ppm |
NO2 – Nitrogen dioxid |
0 … 1,000ppm |
SO2 – Sulfur dioxide |
0 … 4,000ppm |
|
|
Dimensions and other data |
Display |
3.5” TFT color and backlit display |
Weight / Protection class |
55lbs (25kg) / IP54 |
Dimensions |
27.55″ x 23.61″ x 8.26″ (H x W x D) for wall or rack mounting |