SWG 100 SYNGAS EX- Analyzer
- Designed for use in industrial safe or hazardous area Zone 2, or areas with increased safety levels
- Can be installed in outdoor or indoor locations
- Capable of sampling dry or wet syngas
- Supports pressurized or low-pressure gas sampling
- Usable with single or double sampling points
For use in industrial safe or hazardous areas Zone 2, or in areas with an increased safety level.
Basic Equipment
- The stationary analyzer for continuous Syngas measurements in ZONE 2
- For use in industrial safe or hazardous areas Zone 2, or areas with increased safety levels
- Installation suitable for outdoor or indoor locations
- Sampling of dry or wet syngas for high or low-pressure applications
- Low-pressurized gas usable for multiple sampling points
- IP65 stainless steel cabinet suitable for outdoor use with sun and rain protection
- All stainless steel internal piping and fittings
- Accurate measurements using NDIR and thermal conductivity technology
- Sampling range from -100 mbar to +200 mbar gas pressure
- Direct measurement without dilution with pressure and temperature compensation
- Up to 6 gas sampling ports plus 1 calibration gas port
- Optional configuration of up to 3 additional calibration gas ports
- Flow restriction and automatic gas cut-off in case of alarm
- Common process gas recirculation for all sample inputs
- Industry-compatible rugged and service-friendly design
- Ready-to-run delivery with minimal installation effort
OPTIONALLY
- NDIR bench for CO, CO₂, and CH₄ analysis (selective CH₄ measurement)
- O₂ paramagnetic cell measurement
- H₂S electrochemical cell measurement (H₂ immune)
- H₂ thermal conductivity detector with cross-interference correction
- RS485 to USB, Profibus, or Ethernet converter
- I/O modules with 4-channel 4–20 mA analog outputs and 2 alarm relays
- Sample gas washing device for tar-free gas
- Certified ZONE 2 extraction probes
- Heated and temperature-regulated sampling lines in various lengths
1. Does it need to be installed outside of the hazardous (C1D2) area?
Yes, the standard configuration does not meet C1D2 classification requirements. However, the SWG-100 can be outfitted for C1D2 conditions with an additional stainless steel enclosure and constant purge system requiring compressed air.
2. Does it need to be installed under a canopy or shaded area?
If installed outdoors, it should be placed under a canopy or shaded area.
3. What is recommended for sample lines?
PTFE gas sampling lines (DN4/6mm) up to 200 meters are recommended. Stainless steel tubing is also acceptable. Heat tracing is recommended in cold environments.
4. Can you quote a 115VAC power supply?
Yes, 115VAC can be supplied with no price change.
5. Does it require calibration or carrier gas?
Calibration gas is optional unless auto-calibration is required.